48
NeonateBody
Indications:
Spiral mode for routine neonate body studies, e. g.
tumors, abnormalities, malformations, abscesses, etc.
NeonateBody
kV 80
Effective mAs 25
Slice collimation 1.5 mm
Slice width 6.0 mm
Feed/Rotation 24 mm
Rotation time 0.5 sec.
Kernel B30f
Increment 6.0 mm
Image order cr-ca
CTDI
w
0.55 mGy
Effective dose* Male: 0.5 mSv
Female: 0.7 mSv
Tips
You can modify the slice width for image reconstruc-
tion according to the clinical indications.
Contrast medium IV injection
Start delay exam dependent
Flow rate dependent upon needle size/Access site
Total amount 1 2 ml per kg of body weight
* The conversion factor for an 8-week-old,
and a scan range of 150 mm was used.