ExtremityUHRSeq
Indications:
Sequence mode for Ultra High Resolution bone studies,
e. g. tumors, post-traumatic changes, orthopedic indi-
cations, etc.
ExtremityUHRSeq
kV 120
Effective mAs *
Slice collimation 0.6 mm
Slice width 0.6 mm
Feed/Scan 1.0 mm
Rotation time 0.75 sec.
Kernel U80u
Image order cr-ca
* The mAs should be adjusted to the body weight
of the child.
Children
Body weight kV mAs CTDI
w
(mGy)
< 15 kg 120 26 3.3
15 34 kg 120 40 5.1
35 54 kg 120 80 10.3