Page 22 of 22 7482255 Rev 01 Siemens
October 2002
Diagnostic Flow Charts
Figure 6-12 Chart H: DIMAQ-IP
H
IPPC display
Ok?
CINE clip
function Ok?
All IPPC
problems
resolved.
Loop test 1
Ok?
Loop test 2
Ok?
Return
Replace IPPC module.
Replace host module.
Replace cable.
Replace IPPC module.
Replace IPPC module or host
module.
Perform loop test 1.
Service >
Diagnostics>
IPPC Pattern Check.
Perform loop test 2.
Service >
Diagnostics>
IPPC Pattern Check.
No
No
No
Yes
Yes
Yes
Yes
Yes
No
No