Page 4 of 42 7482255 Rev 01 Siemens
October 2002
Installing or Updating the System Software
4. If the DIMAQ-IP software will be updated, back up all exam and patient
data to a CD before performing the DIMAQ-IP software update.
5. If deemed necessary, uninstall the feature keys from the system according
to the instructions provided in "Uninstalling the Product Feature Keys"
(page 8-14).
6. If the system software is being updated to version 1.2, update the DIMAQ-
IP software to version 1.2 before continuing this procedure: See "Install-
ing or Updating the DIMAQ-IP Software" (page 8-7).
7. Insert the MO diskette containing the version 1.2 software in the MO drive.
8. Press the F4 key to display the Preset screen shown in Figure 8-1.
!
Caution: If the system software will be upgraded (or downgraded to an earlier
version) for testing or to resolve an operational problem, the feature keys for
options that are not supported by the installed version will be lost. To avoid this
problem, uninstall all feature keys before upgrading or downgrading the system
software. See "Uninstalling the Product Feature Keys" (page 8-14).
The uninstallation of all feature keys is a precaution. Experienced customer ser-
vice engineers (CSEs) may choose to leave the keys installed.
Note: The version 1.2 software update is described in "Installing the Version 1.2
Software Update on the G60 S" (page 7-1)
.
Figure 8-1 Preset Screen