Page 8 of 42 7482255 Rev 01 Siemens
October 2002
Installing or Updating the DIMAQ-IP Software
1. Back up all exam and patient data to a CD before continuing with the
DIMAQ-IP software update.
2. On the G60 S, save the feature key for the Stress Echo option by perform-
ing the procedure "Uninstalling the Product Feature Keys" (page 8-14).
3. Open the upper-rear panel of the system to expose the Microsoft Windows
label attached to the DIMAQ-IP.
4. Record the 20 characters of the Microsoft Windows 2000 Product Identifi-
cation code; this license code is required to enable the new software.
5. Close the upper-rear panel.
6. Press the F4 key and select the Service menu item.
7. Access the Service software according to the instructions in "Service Soft-
ware Operation" (page 8-17).
8. Select the DIMAQ-IP Maintenance button located on the Main Menu
screen (Figure 8-10) to display the Windows 2000 screen running on the
DIMAQ-IP module.
9. Insert the CD containing the version 1.2 DIMAQ-IP software, in the CD
drive.
10. Move the mouse pointer to the bottom of the screen to display the Windows
toolbar, and select the Start button.
11. Select Shutdown
Restart from the menu.
12. The DIMAQ-IP will reboot from the software CD; the display of the BIOS
and Windows boot screens will be followed by the software installation win-
dow.
13. The blue Radisys logo will be displayed.
14. The software installation requires approximately 5 minutes: The R:\>
prompt appears when the installation is completed.
15. After the R:\> prompt appears, remove the software CD.
16. Operate the power switch to shut down the system: After two minutes, if the
system has not powered down, operate the circuit breaker located at the
rear of the system to de-energize it.
17. After the system has powered down, wait 15 seconds and re-energize the
system.
!
Caution: Failure of the DIMAQ-IP software installation or update will not nor-
mally destroy presets or service data. As a precaution, this data can be backed
up according to the procedures provided in "Backup/Restore" (page 8-36) and
at step 3. (backing up presets) on page 8-2.
Note: The Stress Echo feature key must be reinstalled after the DIMAQ-IP soft-
ware is updated. See "Installing the Product Feature Keys" (page 8-10).
Note: The DIMAQ-IP will boot from the CD drive if Windows is restarted with a
bootable CD installed in the CD drive.