Siemens 7482255 Rev 01 Page 39 of 42
October 2002
Chapter 8 - Software
7. When the progress bar located on the Backup / Restore screen reaches
100% a Done message will be displayed: Remove the MO disk and label it.
6.11.2 Restoring Service Data
1. Select the Restore radio button located at the top of the screen; in the
scrolling list, check the box adjacent to the service data items to be
restored.
2. Insert the MO disk that contains the appropriate backup into the MO drive.
3. Select the Go button located at the bottom of the screen to begin the
restore process.
The restoration process does not allow the user to select individual packages
for restoration. The broader categories available for restoration are listed in
Table 8-7.
6.12 Imaging Test Pattern
Access the Imaging Test Pattern screen shown in Figure 8-28 by selecting the
Imaging Test Pattern button located on the Main Menu screen (Figure 8-10).
To display a test image, select the appropriate button located on this screen.
When a test image is displayed, select the Next key to return to the Imaging
Test Pattern screen.
Adjust the VCR, VTR, and the monitor while displaying the test images
accessed from this screen:
Black and White Test Pattern: See Figure 8-29
Framing and Black and White Grayscale Pattern: See Figure 8-30
Color Bar Pattern: See Figure 8-31
!
Caution: Selecting the Cancel button during an incomplete service data restore
procedure will leave the system inoperable: Do no select the Cancel button
when restoring service data.
Table 8-7 Service Data Restore Hierarchy
Top Level Individual Packages Comment
Service All service data
Site Information Site only
Peripheral Information Peripheral only
Service History Service History only
Note: Terminate the display of a test pattern by pressing SET.