Installing the G50 or the G60 S
2.0 Installing the G50 or the G60 S
Use the following instructions to install the ultrasound system at the customer
site.
2.1 Tools Required
No special tools are required for this installation; only standard service tools
are needed.
2.2 Documents Required
The Field Electrical Safety Test Procedure (p/n 7482776), the G50 and G60 S
Electrical Safety Test Limits (p/n 7482784), and the Field Electrical Safety Test
Report (p/n 7482768) are used during the installation procedure.
2.3 Preparation
Read this entire procedure before beginning an installation. Be prepared to
submit an installation report as described in "Installation Report" (page B-1).
2.4 Installation
1. Unpack the system according to the guidelines provided in "Unpacking the
G50 and the G60 S" (page A-2).
2. Test and verify the operation of the caster locking mechanism on all casters
and the lock mechanism on the monitor arm.
3. Install the transducer and gel bottle holders.
4. Perform the system safety test according to "Safety Checks" (page 9-5).
5. Perform the Field Electrical Safety Test Procedure (p/n 7482776) using the
G50 and G60 S Electrical Safety Test Limits (p/n 7482784); record the
results of the safety test in the Field Electrical Safety Test Report (p/n
7482768).
6. After completion of the electrical safety test and the test report, provide the
customer and the regional office with a copy of the report; retain a copy of
the report for your personal records.
7. Perform the system operational test according to "System Test" (page 9-6).
8. After successful completion of the system test, submit an installation report
according to the instructions provided in "Installation Report" (page B-1).
Note: Advise the customer to store the factory packaging materials for later use.