Removing and Installing the DIMAQ-IP Module
5.0 Removing and Installing the DIMAQ-IP
Module
Use the instructions provided in this section to remove and install the DIMAQ-
IP module on the G50 or the G60 S. The DIMAQ-IP module has also been
referred to as the image processor PC (IPPC). Replacement of the DIMAQ-IP
module requires approximately 15 minutes.
5.1 Tools Required
Use a short #2 Phillips head to remove and install the DIMAQ-IP module.
5.2 Removing the DIMAQ-IP Module
1. Turn off the system and unplug the power cable from wall outlet.
2. Remove the screw caps and the two screws and washers from the top
edge of the rear cover.
3. Loosen two screws on the host module cover, located near the power cord
connector.
4. Remove the host module cover from G50 or the G60 S as you pull out the
top edge of the rear cover; the removal of the cover exposes the view
shown in Figure 4-20.
5. Disconnect all cables from the top surface of the host module (Figure 4-
20).
Note: See "Installing the G50 DIMAQ-IP" (page 5-2) for information on installing
the DIMAQ-IP option on the G50.
!
Warning: The G50 and the G60 S contain hazardous voltages that can range from 90 Vac to
264 Vac. These voltages can cause serious injury or death. Only authorized service personnel
are permitted to remove protective covers and access the energized components of this sys-
tem.