Siemens 7482255 Rev 01 Page 37 of 52
October 2002
Chapter 4 - Removing and Replacing Spare Parts
7. Release the cables from the cable clamp located at the upper left of the
DIMAQ-IP module.
8. Disconnect the remaining cables from the DIMAQ-IP module.
9. Loosen the three thumbscrews of the DIMAQ-IP module (Figure 4-38),
using the short number 2 Phillips screwdriver.
10. Create sufficient clearance for dismounting the DIMAQ-IP module safely by
routing the cables in front of it, and to the sides of the DIMAQ-IP module.
11. Move the DIMAQ-IP module to the left and dismount it from the system.
12. Remove the two screws located under the rear handle and remove the
monitor post cover from upper G50 or the G60 S as shown in Figure 4-32.
13. Loosen the two screws indicated in Figure 4-33.
Figure 4-30 DIMAQ-IP Module Installation
Tie-wrap these cables to the adjacent panel.
DIMAQ-IP
Figure 4-31 DIMAQ-IP Module Thumbscrews
DIMAQ-IP