Siemens 7482255 Rev 01 Page 45 of 52
October 2002
Chapter 4 - Removing and Replacing Spare Parts
10. Remove the EMI shielding bracket.
11. Disconnect the four cables from these connectors of the E module: A66J4,
A66JI, A66J3, and A25J1.
12. Loosen the six screws (three on each side) from the sides of the E module.
13. Loosen the two screws located at the bottom of the front cover, and pull out
the top edge of the cover to unfasten three pairs of magic fasteners joining
the front cover to the system. Hold the bumper, and remove the entire front
cover from the G50 or the G60 S.
14. Loosen the two screws that hold the ECG connector bracket.
15. Remove the six screws (three each side) from the sides of the E module.
16. As indicated in Figure 4-41, pull the plastic loop to move the E module
toward you, while holding the handle located on the top of the E module:
Pull the E module slowly toward you as indicated in Figure 4-41.
Figure 4-40 Removing the EMI Shielding Bracket
!
Caution: Loose hardware can drop into the E module. Loosen all hardware then
remove it carefully. Recover all dropped hardware before continuing.
!
Caution: The removal of large and heavy subsystems can pinch and damage
cables. Verify that system cables are not pinched when the E module is
removed from the system.
Figure 4-41 Removing the E module