Page 46 of 52 7482255 Rev 01 Siemens
October 2002
Removing and Installing the E Module
10.3 Installing the E Module
Install the E module by performing, in reverse order, the procedures provided in
"Removing the E Module" (page 4-44).
10.4 Confirming the Operation of the E Module
1. Connect the system power cable to a properly grounded outlet.
2. Connect a transducer to the G50 or the G60 S.
3. Turn the power switch On.
4. Turn the Standby switch On.
5. Confirm that a B-mode image is displayed on the monitor.
!
Caution: Use fasteners of the correct length to prevent damage to underlying
equipment. See "Fastener Reference" (page H-1).