Page 50 of 52 7482255 Rev 01 Siemens
October 2002
Replacing the OEM Outlet Fuse
12.0 Replacing the OEM Outlet Fuse
Use the instructions provided in this section to remove and install the power
fuse located in the host module. Replacement of the fuse requires
approximately 5 minutes.
12.1 Tools Required
Use a number 2 Phillips screwdriver to perform this procedure.
12.2 Replacing the Fuses
1. Turn off the system and unplug the power cable from the wall outlet.
2. Loosen the two captive screws from the upper side of the rear cover and
remove the cover.
3. Loosen the two captive screws from the lower rear panel and remove the
panel.
4. Pull the upper side of the rear cover toward you.
5. Locate the XDST (A07-1) PC board on the right side of the host module;
see Figure 4-44.
!
Caution: An electrical failure within the system causes the power fuse(s) to
open. Before replacing the power fuse, discover and correct the failure that
caused the fuse to open.
!
Warning: The G50 and the G60 S contain hazardous voltages that can range from 90 Vac to
264 Vac. These voltages can cause serious injury or death. Only authorized service personnel
are permitted to remove protective covers and access the energized components of this sys-
tem.
Figure 4-44 XDST A07-1 (07473122)