Technical Description
3-4
Operating instructions ULTRAMAT 23 gas analyzer
C79000-B5276-C216-03
Executions of the wetted parts
Gas path 19“ unit Desktop unit
With hoses Condensation trap/gas inlet - PA (polyamide)
Condensation trap - PE (polyethylene)
Gas nipples 6 mm PA (polyamide) PA (polyamide)
Gas nipples ¼“ SS, type No. 1.4571 SS, type No. 1.4571
Hose FPM (Viton) FPM (Viton)
Pressure switch FPM (Viton) + PA6-3-T (Trogamide) FPM (Viton) + PA6-3-T (Trogamide)
Flowmeter PDM/Duran glass/X10CrNiTi1810 PDM/Duran glass/X10CrNiTi1810
Elbows/T-pieces PA6 PA6
Internal pump PVDF/PTFE/EPDM/FPM/Trolene/ SS, type
No. 1.4571
PVDF/PTFE/EPDM/FPM/Trolene/ SS, type
No. 1.4571
Solenoid valve FPM70/Ultramide/1.4310/1.4305 FPM70/Ultramide/1.4310/1.4305
Safety reservoir PA66/NBR/PA6 PA66/NBR/PA6
Sample cell
Body Aluminium Aluminium
Cell lining Aluminium Aluminium
Stub SS, type No. 1.4571 SS, type No. 1.4571
Window CaF2 CaF2
Adhesive E353 E353
O-ring FPM (Viton) FPM (Viton)
With pipes Gas nipples (6 mm / ¼ “) SS, type No. 1.4571
Pipes SS, type No. 1.4571
Sample cell
Body Aluminium
Cell lining Aluminium
Stub SS, type No. 1.4571
•Window CaF2
Adhesive E353
O-ring FPM (Viton)