Technical Description
Operating instructions ULTRAMAT 23 gas analyzer
C79000-B5276-C216-03
3-13
3.5 Technical Data
General
Measured components Max. 4, of which up to 3 infrared-
sensitive gases plus oxygen
Measuring ranges 2 per channel
Characteristics Linearized
Conformity CE-identification to EN 61326/A1,
EN 61010-1
Display LCD with LED backlighting and
contrast control, function keys
80 characters
(4 lines/20 characters)
Position of use Front panel vertical
Design, enclosure
Weight Approx. 10 kg
Degree of protection, 19" unit and
desktop unit
IP20 to EN 60529
Electrical characteristics
EMC interference immunity (Electro-
Magnetic Compatibility), safety
extra-low voltage (SELV) with safe
isolation
According to standard require-
ments of NAMUR NE21 (08/98) or
EN 50081-1, EN 50082-2
Power supply 100 V AC, +10%/-15%, 50 Hz,
120 V AC, +10%/-15%, 50 Hz,
200 V AC, +10%/-15%, 50 Hz,
230 V AC, +10%/-15%, 50 Hz,
100 V AC, +10%/-15%, 60 Hz,
120 V AC, +10%/-15%, 60 Hz,
230 V AC, +10%/-15%, 60 Hz
Power consumption Approx. 60 VA
Electric inputs and outputs
Analog output per component, 0/2/4 to 20 mA,
floating, max. load 750 Ω
Relay outputs 8, with changeover contacts,
freely selectable, e.g. for range
identification, loading capacity,
24 V AC/DC /1 A, floating, non
sparking
Binary inputs 3, designed for 24 V, floating
•Pump
AUTOCAL
Synchronization
Serial interface RS 485
AUTOCAL function Automatic analyzer calibration
with ambient air (depending on
measured component), cycle
time adjustable from 0 (1) to
24 hours
Options Supplementary electronics with 8
additional binary inputs and relay
outputs, e.g. for external auto-
matic calibration and for PROFI-
BUS PA or PROFIBUS DP
Ambient conditions
Permissible ambient temperature
Operation +5 to +45 °C
Storage and transport -20 to +60 °C
Permissible ambient humidity < 90% RH (relative humidity) for
storage and transport
Permissible pressure variations 700 to 1200 hPa
Sample gas
Sample gas pressure
Without pump unpressurized
With pump unpressurized suction mode, fac-
tory-adjusted with 2-m hose at
sample gas outlet; an upper
range value calibration is
required in case of other venting
conditions
Sample gas flow 72 to120 l/h (1.1 to 2 l/min)
Sample gas temperature 0 to 50 °C
Sample gas humidity < 90% RH (relative humidity)
no condensation