Functions
8.1 Diagnostics
ULTRAMAT 23
104 Manual, 3/2016, A5E37100388-003
In this menu you can call all status messages of the
ULTRAMAT
23 via further submenu items.
tatus of the O
2
sensor is shown in the last
line. If the analyzer is equipped with software for operating an
H
2
S sensor, the text 'Probe status' is shown in the last line with
Only H
2
S sensor:
The analyzer calls this function directly.
H
2
S and O
2
sensors:
You will be requested to select the associated sensor.
Analyzer status: Status: Logbook/faults
This dialog displays the contents of a l
ogbook. This con-
tains all recorded faults. Each type of fault only appears
once in the logbook and is output in alphanumeric text (an
overview of the possible faults can be found in section
Faults
(Page 178)).
Viewing the logbook is protected by code level
1.
Following access to the logbook you can:
Display all recorded faults in succession using the
key
Delete the currently displayed fault message using the
key.
If further faults are present, these appear in succession.
You should delete all fault messages whose causes
have been eliminated.
By deleting the fault message you do no
t eliminate the
cause of the fault (see section
Faults (Page 178)).
A corresponding text is output when all stored faults have
been displayed. Te
rminate display of the logbook using the
key.
If the analyzer is in measuring mode, you can recognize
the occurrence of a fault in that an "F
" appears at the right
at the right edge signals that a fault has been
logged which is no longer present.