Functions
8.1 Diagnostics
ULTRAMAT 23
Manual, 3/2016, A5E37100388-003
107
8.1.2
Diagnostics: Diagnostics values
The diagnostics values provide important information for tro
u-
bleshooting and adjustments.
You can select the four dis-
played function groups in this menu.
If the analyzer contains software for H
2
S measurement, the
adjacent di
splay appears. The following versions are possible
following selection of the 'Probe diagnostic values' parameter
in the 2nd line:
Only H
2
S sensor: Following selection of this item, a branch
is made to the diagnostics values of the H
2
S sensor
(section
Analyzer status: Diagnostics values: H2S sensor
(Page 109)).
H
2
S and O
2
sensors: A query is made for the component,
and a branch then made to the called sensor.
8.1.2.1
Analyzer status: Diagnostics values: IR
In this submenu you can call the diagnostics values of the i
n-
frared measuring ranges. These are:
ADU
are voltage and signal values of the analog-to-digital
converter prior to temperature compensation. These values
are specific to the component.
V-ADUt
are voltage and signal values of the analog-to-
digital converter following temperature compensation.
T
in the bottom line corresponds to the temperature of the
analyzer unit,
TS
the temperature of the IR source
(empty field = function not yet implemented,
"*****" = no measured value present).