Functions
8.1 Diagnostics
ULTRAMAT 23
Manual, 3/2016, A5E37100388-003
109
Analyzer status: Diagnostics values: H2S sensor
diagnostics values of the optional H
2
S
. is the actual current of the H
2
S sensor in nA
2
is the current H
2
S measured value in vpm. Negative
values are also possible here.
is the installation date of the H
2
S sensor (see
section
Calibration: H2S sensor: Defining the installation
(Page 119))
is the pressure-compensated current per vpm H
2
S
of the sensor when it was installed.
Diagnostics: Diagnostics values: Pressure sensor
This dialog displays the diagnostics values of the pre
s-
Calibration: Pressure sensor
121)). The displayed values have the following
is the actual voltage of the pressure
sensor measured at the output of the A/D converter.
is the actual atmospheric pressure in
mbar.
Analyzer status: Diagnostics values: Other diagnostics values
This menu is used to call further diagnostics functions. You
can call the following values:
–
: Data on the supply voltage in % of the nominal
value of the respective supply voltage (e.g.: 100%
corresponds to 230 V or 120 V).
–
: This is the raw voltage following the
rectification
–
,
: Data on the IR source voltage(s), in
Volt. An empty value indicates that the corresponding
source does not exist.