Functions
8.3 Parameter
ULTRAMAT 23
128 Manual, 3/2016, A5E37100388-003
Two limits are assigned to eac
h component, and can be set
using this menu. A relay is triggered when the limits are viola
t-
Configuration: Inputs/outputs/pump: Assign
(Page 138)). Limit 1 is the lower limit, Limit 2 the upper
Note
A set limit only triggers a relay contact if a relay has previously been assigned to the
corresponding limit signal (see section
Configuration: Inputs/outputs/pump: Assign relays
138)). The limits are not updated:
During the first warm-up phase
During an AUTOCAL
During the message: Function control and analog output at 'Hold measured value' (see
section
Configuration: Inputs/outputs/pump: Analog outputs (Page 134))
If you have selected "Limit 1" or "Limit 2", a menu display a
p-
pears in which you can enter the lower or upper limit for each
component. You can define the
value in the second line, and
the condition under which a contact is be triggered in the third
(Alarm at conc.:):
: with upward violation
: with downward violation
: no signal.