Functions
8.4 Configuration
ULTRAMAT 23
Manual, 3/2016, A5E37100388-003
139
Table 8- 4 Overview of relay assignments
Limit Limit has been triggered Limit
(see section Parameters: Limits
Measuring range Range 2 Range 1 -
Maintenance request Maintenance
Function control Function control
present
During warm-up phase (approx. 30
min), AUTOCAL,
Calibration gas (sample
Calibration gas flowing Ext. solenoid valve open
Sync. Synchronization signal is
output
AUTOCAL only "Zero gas flow" and
adjustment not during sample gas
The pin assignments of the relays in the de-energized state are described in section Pin
assignments (Page 54).
The functions which can be assigned to the relays have the following meanings:
The upper and lower limits can be defined as events for tri
g-
ring relays. Select the desired relay(s) in the second and
third lines of this menu.
This function is specific to the component.
A relay for range switchover can be assigned to each comp
o-
nent. This guarantees reliable a
ssignment of the analog output
signal to the currently active range, especially with autoranging
(see section Parameter: Measuring ranges (Page 125)).