8.4 Configuration
ULTRAMAT 23
Manual, 3/2016, A5E37100388-003
139
Table 8- 4 Overview of relay assignments
Function
Relay de-energized
Relay energized
Signaling
Limit Limit has been triggered Limit
(see section Parameters: Limits
(Page 128))
Measuring range Range 2 Range 1 -
Status messages
Fault
Fault present
Maintenance request Maintenance
request
CAL/MEAS
Measure
AUTOCAL
AUTOCAL
Function control Function control
present
During warm-up phase (approx. 30
min), AUTOCAL,
uncoded
External solenoid valves
Zero gas
Zero gas flowing
Ext. solenoid valve open
Calibration gas (sample
gas inlet)
Calibration gas flowing Ext. solenoid valve open
Sync. Synchronization signal is
output
AUTOCAL only "Zero gas flow" and
adjustment not during sample gas
purging phase
The pin assignments of the relays in the de-energized state are described in section Pin
assignments (Page 54).
The functions which can be assigned to the relays have the following meanings:
Limit messages
The upper and lower limits can be defined as events for tri
g-
ge
ring relays. Select the desired relay(s) in the second and
third lines of this menu.
This function is specific to the component.
Relays for MR
A relay for range switchover can be assigned to each comp
o-
nent. This guarantees reliable a
ssignment of the analog output
signal to the currently active range, especially with autoranging
(see section Parameter: Measuring ranges (Page 125)).