Functions
8.4 Configuration
ULTRAMAT 23
Manual, 3/2016, A5E37100388-003
141
Configuration: Inputs/outputs/pump: Binary/sync inputs
You can use this dialog to set the response of the synchroniz
a-
tion input and the binary inputs. Select one of the adjacent
options:
You can use this dialog to set the response of the synchroniz
a-
tion input. This function allows an AUTO
CAL procedure to be
triggered simultaneously for several analyzers within a system.
You can select between the following settings in the third line
(see also section
System setup with several analyzers in paral-
(Page 84)):
:
The analyzer carries out an AUTOCAL and activates its
sync output up to the end of the electronic adjustment.
If the flow becomes too low during a zero adjustment
tr
iggered via the sync input, this zero adjustment is aborted,
and an error status set. This aborted zero adjustment is
entered in the logbook.
(set in Fig.):
The analyzer enters the CAL status, but does not carry out
an AUTOCAL. The analyzer waits until the Sync input
becomes inactive. It then enters the status 'Purge sample
gas' and subsequently selects measuring mode.
this dialog to freely configure 8 floating binary
V (0...4.5 V); "1" = 24 V (13...33
V)] in analyzers
with an add
-on board. The pin assignments of the 37-contact
plug are described in section
Pin assignments (Page 54). No
inputs are preset on delivery.
The adjacent error message occu
rs if an attempt is made to
call this function for a device without add-on board.
The adjacent display appears if an add
-on board is present.
the following functions to the eight binary
Seven different messages for faults/maintenance requests
Four different messages for function control
Switch ranges
Delete the logbook.