Error and system messages
11.2 Faults
ULTRAMAT 23
180 Manual, 3/2016, A5E37100388-003
AUTOCAL drift beyond tolerance Detector contaminated
IR source power too low
EEPROM error
Read character does not correspond to
Channel 1 not calibrated Calibration of full-scale value / sag miss-
Channel 2 not calibrated Calibration of full-scale value / sag miss-
Channel 3 not calibrated Calibration of full-scale value / sag miss-
Voltage for IR source beyond toler-
ance
Bridge supply voltage outside toler-
ance
Half-bridge voltage outside tolerance
Lockin error
Sensitivity of O
2
sensor too low O
2
sensor faulty or no longer usable as
result of aging
Replace the O
2
sensor as described in
section
Replacing the electrochemical
External fault Signaling of an external fault (system-
specific)
Check the connected devices for faults
as described in section Configuration:
Inputs/outputs/pump: Binary/sync inputs
H
2
S protection function Sample gas concentration too high Check sample gas, see also section
Probe protection function (Page 156)
Zero point of the H
2
S sensor beyond
Purge time too short for calibration Repeat calibration
Sensitivity of H
2
S sensor too low