3.4 Technical specifications
ULTRAMAT 23
36 Manual, 3/2016, A5E37100388-003
Influencing variables
Drift
With AUTOCAL
Without AUTOCAL
Negligible
< 2 % of smallest measuring range/week
Temperature max. 2 % of smallest possible measuring range according to label per
10 K with an AUTOCAL cycle time of 6 h
Air pressure
&lt; 0.2 % of measuring range per 1 % change in pressure
Power supply
&lt; 0.1 % of output signal span with a variation of ±10 %
Line frequency ± 2 % of full-scale value with a frequency variation
of ± 5 %
Deviations with measuring range 0 ... 200 mg/m
3
SO
2
(analyzer versions 7MB2335-xNBxx-xAAx, 7MB2337-xNBxx, 7MB2337-xxxxx-xNBx, 7MB2338-xxxxx-xNBx)
Availability
Max. 95 %
AUTOCAL cycle time
Max. 6 h
Temperature variations Max. 1 °C (1.8 °F)
The device must not be operated in an area subject to drafts. This is
especially valid for the rear panel with large cooling element.
Other
This measuring range has not been suitability-tested.
Deviations with measuring range 0 ... 100 mg/m
3
SO
2
(analyzer versions 7MB2335-xNTxx-xAAx, 7MB2337-xNTxx-xxxx, 7MB2337-xxxxx-xNTx, 7MB2338-xxxxx-xNTx)
Delayed display (T
90
time) Depending on the external gas conditioning.
Note: SO
2
is highly soluble in water
Output signal resolution
≤ 0.1 mg/m
3
Linearity error
≤ 2 mg/m³ (2% of smallest measuring range)
Drift
With constant ambient conditions
After temperature jump
≤ 0.33 mg/m³ per hour after a running-in period of 200 hours
Not defined
Temperature influence
≤ 8 mg/m³ per 10 K deviation; additional drift after temperature jump
Influence of auxiliary power 0.4 mg/m³ with a change of ± 10%
Influence of line frequency 8 mg/m³ with a change of ± 5%
Cross-sensitivities
Refer to following table
Availability
Max. 95 %
AUTOCAL cycle time
Max. 3 h
Temperature variations Max. 1 °C (1.8 °F)
The device must be operated at a constant ambient temperature. The
device must not be operated in an area subject to drafts. This is especial-
ly valid for the rear panel with large cooling element.
Other
This measuring range has not been suitability-tested.