Description
3.4 Technical specifications
ULTRAMAT 23
40 Manual, 3/2016, A5E37100388-003
Permissible ambient temperature
• During operation
• During transportation and storage
+5 ... +45 °C (41 ... 113 °F)
-20 ... +60 °C (-4 ... 140 °F)
Permissible ambient humidity
< 90% RH (relative humidity) during transportation and storage
Permissible ambient pressure
600 ... 1200 hPa absolute
Maximum operating altitude
Cross-inferences (interfering gases)
See table of cross-sensitivities
Zero drift (vol % O
2
) Measuring range 2%: max. 0.1% with weekly zero adjustment
Measuring range 5%: max. 0.1% with weekly zero adjustment
Measuring range 25% or greater: max. 0.5% with monthly zero adjust-
Temperature error (vol % O
2
) <2% /10 K referred to measuring range 5%
<5% /10 K referred to measuring range 2%
Humidity error (vol % O
2
)
2
with 90% relative humidity after 30 min
<0.6% at 50 °C (122 °F)
< 0.2% of measured value per 1% pressure change
Cross-sensitivities
All values in this table refer to a zero calibration with nitrogen and a full-scale calibration with 100 vol. % oxygen. The devi-
ations apply to 100 vol. % of the associated gas, and must be considered proportionally for the zero calibration.
2
4
3
6
2
2
3
6
6
2
4
6
4
10
4
8
Chlorine Cl
2
- 0.83 - 0.91
Diacetylene C
4
H
2
- 1.09 - 1.20