Make sure when mounting the analyzer that the environment is as free as possible of the
gas components to be measured!
In order to achieve the highest possible measuring quality, also observe the following
information concerning the location for mounting an analyzer!
The device may overheat or start burning in the case of insufficient ventilation.
• Ensure sufficient ventilation between the devices when installing in control cabinets. The
heat sinks at the rear must remain free for air circulation.
• Make sure during operation that the permissible ambient temperature range is always
observed (see Technical specifications (Page 31)).
The device can be damaged, destroyed, or its functionality impaired through improper
mounting.
• Before installing ensure there is no visible damage to the device.
• Make sure that process connectors are clean, and suitable gaskets and glands are
used.
• Mount the device using suitable tools. Refer to the information in Technical
specifications (Page 31) for installation torque requirements.
Strong vibrations could loosen connections or damage sensors, resulting in free passage of
the sample gas into the environment.
Even weaker vibrations influence the result!
The analyzer must therefore only be used at a location which is free of vibration.
Please observe the data in the Technical specifications. (Page 31)