Commissioning
6.5 System setup with several analyzers in parallel
ULTRAMAT 23
Manual, 3/2016, A5E37100388-003
87
Slave:
Set the AUTOCAL cycle time to "0" to prevent a cyclic AUTOCAL from being triggered
(see Calibration: AUTOCAL/drift values: Cycle time (Page 123)).
Assign the "Sync." function to a relay (see Configuration: Inputs/outputs/pump: Assign
relays (Page 138)).
Assign the "AUTOCAL" function to the digital input SYNC (see Configuration:
Inputs/outputs/pump: Binary/sync inputs (Page 141)).