3
WarningsandCautions
Warning
•
Danger — Explosion hazard. Do not use in the presence of flammable
anesthetics.
•
Topreventfireorshockhazards,donotimmersetheunitinwaterorexposeitto
rainormoisture.
•
Donotusetheunitwithanextensioncordreceptacleorotheroutletsunlessthe
prongsofthepowercordcanbefullyinserted.
•
RISK OF ELECTRICAL SHOCK — DO NOT OPEN. To reduce the risk of electrical
shock, DO NOT remove the back of the equipment or open the enclosure. No
user‐serviceablepartsareinside.Referservicingtoqualifiedfieldserviceengineers
only.
•
Uninsulated voltage within the unit may have sufficient magnitude to cause
electricalshock.Avoidcontactwithanypartinsidetheunit.
Caution
• BeforeconnectingthecablestoyourEloTo u chSystemstouchmonitor,makesure
allcomponentsarepoweredOFF.
• The use of ACCESSORY equipment not complying with the equivalent safety
requirements of this equipment may lead to a reduced safe ty of the resulting
system. Consideration relating to the choices of accessory equipment
should
include:
• Useofaccessoryinthepatientvicinity.
• Evidencethatthesafetycertificationoftheaccessoryhasbeenperformedin
accordance to the appropriate IEC 60601‐1 and/or IEC 60601‐1‐1
harmonizednationalstandard.
• Forcontinuedsafety—
‐ This unit only complies to the above
standards if used with a medical
gradepowercord.
‐Amedicalgradepoweradaptor,suchastheonespecified,isrequiredfor
useinamedicalapplication.
Note:
•
This symbol alerts the user to important information concerning the operation
andmaintenanceofthisunit,whichshouldbereadcarefullytoavoidproblems.
CAUTION‐LifeSupport
Care must be tak en when this touchmonitor is a critical component of a Life
support system or device. In case of failure of this touchmonitor, appropriate
redundant systems should be incorporated into the system or device to prevent
injurytotheuserorpatient.
The following should be an integral part
of the safety design of a Life support
systemordeviceusingthistouchmonitorforacriticalfunction.
♦An Alternate interfaceorfail‐safemust be available shouldthetouchscreenfail
tooperate.
♦ The touchscreen interface must not be the only means of control of a critical
function.
♦AnAlternatevideomonitorshouldbeincorporatedintothesafetydesignifused