EasyManuals Logo

Smiths Medical CADD-Solis User Manual

Smiths Medical CADD-Solis
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #25 background imageLoading...
Page #25 background image
23
Programming and Operation
WARNING:Topreventthepatientfrom
accessing the clinician bolus function, do
not let the patient know the security codes.
Improper programming could result in
serious patient injury or death.
4. Make sure the clinician
bolus amount is at the
desired value and select
Deliver.
NOTE: If you enter a
value outside the soft
limit, a screen will
appear asking you to
confirm the soft limit
override.
5. The screen will show the
amount decreasing as the
bolus is delivered. You
may stop the bolus at any
time by selecting Stop
Bolus.
WARNING:Neverleavethepumpunattended
while on the clinician bolus edit screen. You
mustpress“Confirm”or“Deliver”todeliver
theprogrammedvalueorcanceltoleavethe
screen. Failure to do so could result in serious
patient injury or death.
Start New Patient
Each time a new patient is started, it will be
recorded in the event log. All other reports will be
cleared.
The pump must be stopped and you’ll need the
keypad code (or a higher level code) or the cassette/
keypad lock must be unlocked, if allowed. To start a
new patient:
1. From the Tasks menu,
press U or D until
Start New Patient is
highlighted and press
S.
a clinician bolus, it will give you the option to restart
the clinician bolus where it left off or to start with a
new clinician bolus.
WARNING:Exercisecarewhenusingtheclinician
bolus function. Since there are no limits on the
frequencyofdeliveringabolus,andsincethe
amount of the bolus can be set as high as 20
mL(orthemgormcgequivalent),youshould
not permit the patient to become familiar with
theprocedureforgivingaclinicianbolus.
Improper programming could result in serious
patient injury or death.
NOTE: The maximum clinician bolus may be limited
by the settings in the protocol, which is determined
by the CADD
®
-Solis system administrator.
To start a clinician bolus:
1. Make sure the pump is running. Start the pump
if necessary.
2. From the Tasks menu,
press U or D until
GiveClinicianBolusis
highlighted, then press
S.
3. Unlock the keypad using
the pump key (if allowed)
or press U or D to
enter the clinician code
(or a higher level code).
Press S to advance to
the next digit. Once the
code has been entered,
select Accept Value.
NOTE: If you enter a
code that you believe
is correct and receive
a wrong code error,
check the screen to see
if the clinician code
has been customized.
If the code has not been customized, it was
entered incorrectly. Retry. If the code has been
customized, use the custom clinician code.
If you do not know this code, contact your
CADD
®
-Solis system administrator.
Press here to Deliver
Press here to Stop Bolus
Press here to Accept Value

Table of Contents

Other manuals for Smiths Medical CADD-Solis

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Smiths Medical CADD-Solis and is the answer not in the manual?

Smiths Medical CADD-Solis Specifications

General IconGeneral
BrandSmiths Medical
ModelCADD-Solis
CategoryMedical Equipment
LanguageEnglish

Related product manuals