6
OPERATOR’S MANUAL
Manufacturer:
Smiths Medical ASD, Inc.
160 Weymouth Street
Rockland, MA 02370, USA
USA/Canada: 1-800-258-5361
International: +1-781-878-8011
www.smiths-medical.com
Made in USA
s s
Level 1
®
NORMOFLO
®
H-1129
Irrigation Fluid Warmer
< H-1129 115V
l
TM
l
TM

Table of Contents

Question and Answer IconNeed help?

Do you have a question about the Smiths Medical Level 1 NORMOFLO H-1129 and is the answer not in the manual?

Smiths Medical Level 1 NORMOFLO H-1129 Specifications

General IconGeneral
ManufacturerSmiths Medical
ModelH-1129
Power Requirements100-240 VAC, 50/60 Hz
Warranty1 year
Product NameLevel 1 NORMOFLO H-1129
TypeFluid Warmer
Temperature Range37°C to 41°C
MaterialPlastic
ColorWhite
Intended UseWarming fluids
DesignPortable, compact

Summary

About this Manual

Conventions Used in this Manual

Explains conventions and symbols used throughout the manual for clarity and consistency.

Description

Gas Vent

Details the function of the Gas Vent in the warming sets, releasing micro-bubbles from fluids.

Temperature Control

Describes the recirculating solution heating system and its temperature control.

Automatically Operating Pole

Explains the electronically operated pole that lifts fluid bags with a touch of a switch.

Level 1® NORMOFLO® Irrigation Warming Sets

Lists the disposable sets available for use with the irrigation fluid warmer.

Indications for Use

Important Safety Information

General Safety Warnings

Covers essential safety information for the safe use of the Fluid Warmer, including electrical and environmental hazards.

Continued Warnings

Provides further critical warnings regarding device usage, accessories, grounding, and operational hazards.

CAUTIONS

Outlines cautions to prevent physical injury or adverse effects on the device or its performance.

Out of the Box-Assembly

Verify Components

Instructions for verifying all components are present before assembly.

Assemble Fluid Warmer to Base

Detailed steps for assembling the main unit to its base.

Install Pressure Chambers

Guidance on installing the pressure chambers onto the device.

Attach Fluid Bag Hanger

Procedure for attaching the fluid bag hanger to the device pole.

Set Up 3-Liter Pressure Accessory

Instructions for setting up the 3-liter pressure accessory.

Disinfect Reservoir

Procedure for disinfecting the internal reservoir before initial use.

Preparation for Use

Final preparation steps before operating the fluid warmer.

Perform Electrical Safety Tests

Instructions for conducting essential electrical safety tests.

Principle of Operation

Fluid Warming

Explains the process of warming fluids using a recirculating solution and heat exchanger.

Pressurized Fluid Delivery

Describes how the warmer uses pressure chambers for fast fluid delivery.

Controls and Displays

Fluid Warmer Display Panel

Details the LCD display and LEDs providing information on temperature and operating modes.

Power and Alarm Test Panel

Describes the buttons for controlling power and testing alarms on the device.

Reservoir Level Display

Explains the window for checking the recirculating solution level in the reservoir.

Pressure Chamber Control Panel

Covers the toggle switch and pressure gauge for managing pressure chambers.

Pressure Gauge Box

Details the pressure dial gauge and regulator knob for pressure adjustments.

Interlocks

Explains interlocks that detect correct installation of disposable sets.

Operation

OFF Mode

Describes how to turn off the fluid warmer and its associated power states.

ON/Automatic Operation Mode

Explains the automatic mode activated when a disposable set is installed and the unit is ON.

Alarm Test Mode

Details how to test the visual and audible alarm indicators of the fluid warmer.

Over Temperature Test Mode

Explains how to test the operation of the fluid warmer's over temperature circuitry.

Check Disposables Mode

Indicates a missing or improperly installed disposable set, stopping circulation.

Add Recirculating Solution Mode

Signifies a low recirculating solution level, stopping circulation and warming.

Over Temperature Alarm Mode

Activates when reservoir solution exceeds 43.9°C, stopping warming and signaling an alarm.

Pressurized Mode

Describes how pressure chambers deliver fluids at increased flow rates.

Unpressurized Mode

Explains how to disable pressurization by moving the toggle switch to the unpressurized position.

Pressure Regulation Mode

Details the pressure regulator knob and dial gauge for adjusting and displaying chamber pressure.

Operating Instructions

General Safety Warnings

Provides crucial warnings regarding device usage, accessories, grounding, and operational hazards.

CAUTIONS

Outlines cautions to prevent physical injury or adverse effects on the device or its performance.

Set Up for Use

Covers setup procedures, including warnings for disposable set installation and usage.

Install Irrigation Warming Set

Step-by-step guide for installing the disposable irrigation warming set onto the fluid warmer.

Prime Irrigating Disposable Sets

Instructions for priming the disposable sets, including warnings and specific set usage.

Attach Disposable Hand Piece

Procedure for connecting disposable hand pieces or specialized irrigation disposables using a standard spike connector.

Test Audible and Visual Alarms

Procedure for testing the visual and audible alarm signals of the fluid warmer.

Use of the Fluid Warmer

Details on pressurizing chambers, preparing accessories, and adjusting the automatic pole.

Change the Fluid Bag

Steps for safely changing the fluid bag while maintaining pressure and operation.

After Use

Procedures for shutdown, cleaning, and disposal after using the fluid warmer.

Troubleshooting

General Troubleshooting Guide

Provides a guide to identify and resolve common problems with the fluid warmer.

Slow Flow Rate Troubleshooting

Offers solutions for issues related to slow or impeded fluid flow during operation.

Testing

Add Recirculating Solution Alarm Test

Procedure to test the low recirculating solution alarm and its activation.

Check Disposables Alarm Test

Tests interlocks that detect proper installation of the disposable set components.

Over Temperature Test

Tests the fluid warmer's over temperature circuitry and alarm activation.

Fluid Warmer Alarm Signal Test

Confirms the proper operation of all visual and audible alarm indicators.

Performance Testing

Evaluates operational performance, including cold start, calibration, and specific tests.

Periodic Electrical Testing

Procedures for testing earth leakage and ground continuity according to safety standards.

Maintenance

Maintenance Performed with Every Use

Cleaning and inspection tasks to be performed after each use of the fluid warmer.

Maintenance Performed Every 30 Days

Tasks scheduled for every 30 days, including lubricating O-rings and changing solution.

Maintenance Performed Every 12 Months

Annual maintenance tasks, including disinfecting the reservoir and replacing O-rings.

Change O-Rings

Instructions for replacing O-rings in the #1 and #2 blocks.

Clean Fan Filter

Procedure for cleaning the fan filter located on the bottom of the fluid warmer.

Maintenance and Calibration Log

Provides a checklist for scheduled maintenance and calibration tasks.

Limited Warranty

Parties Covered by this Warranty

Specifies who is covered by the warranty, excluding subsequent purchasers.

Warranty Performance Procedure

Details the process for claiming warranty service, including obtaining authorization.

Conditions of Warranty

Outlines conditions that will void the warranty, such as misuse or unauthorized repair.

Limitations and Exclusions

Lists limitations on the manufacturer's liability and excludes implied warranties.

Service

WARNING (Service)

Emphasizes that only authorized personnel should perform service on the device.

Warranty Service

Describes the process for obtaining free repair for units under warranty.

Non-Warranty Work

Explains procedures for repair of units that are out of warranty or damaged.

Additional Documentation

Lists documentation available upon request from Smiths Medical.

Disposal Information

Provides guidance on proper disposal of the device, considering hazardous materials.

Service Contacts

Lists contact information for Smiths Medical technical service and distributors.

Specifications

System Specifications

Details product safety, EMC, enclosure protection, physical dimensions, environmental, and electrical specifications.

Electromagnetic Environmental Recommendations

Provides recommended separation distances for RF communications equipment to prevent interference.

Irrigation Warming Sets (Disposable Sets)

Lists the available disposable irrigation warming sets with their product descriptions.

Disposable Set Accessories

Lists available accessories for disposable irrigation sets.

Symbols

Related product manuals