4
This CAUTION is located on the rear of the unit.
See page 16 of these instructions for details about how
to attach the connector cover.
Refer to the operating instructions
Follow the directions in the operating
instructions for parts of the unit on which this
mark appears.
This symbol indicates the manufacturer, and
appears next to the manufacturer’s name and
address.
Important EMC notices for use in the medical environments
The LMD-1530MD needs special precautions
regarding EMC and needs to be installed and put into
service according to the EMC information provided in
this instructions for use.
The portable and mobile RF communications
equipment such as cellular phones can affect the
LMD-1530MD.
Warning
The use of accessories and cables other than those
specified, with the exception of replacement parts sold
by Sony Corporation, may result in increased emissions
or decreased immunity of the LMD-1530MD.
Warning
If the LMD-1530MD should be used adjacent to or
stacked with other equipment, it should be observed to
verify normal operation in the configuration in which
it will be used.
CAUTION
Please provide with the
protection cover for the
connector when you do
not use the specified
connectors.
Guidance and manufacturer’s declaration-electromagnetic emissions
The LMD-1530MD is intended for use in the electromagnetic environment specified below.
The customer or the user of the LMD-1530MD should assure that it is used in such an environment.
Emission test Compliance Electromagnetic environment-guidance
RF emissions
CISPR 11
Group 1
The LMD-1530MD uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The LMD-1530MD is suitable for use in all establishments,
including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class D
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies