34
COMMAND MODULE
STATUS CODES
LED INDICATOR FAULT / WARNING SONAR POWER ACTION FOR SONAR OPERATOR
No Messages from ARIS ON
Cycle power to ARIS (network okay but sonar firmware not running).
No Links with ARIS ON
Check that cable from Command Module (CM) Ethernet connector is attached
to ARIS. If connected, check for dirty or broken contacts at each end.
No Ethernet to PC ON
Check that cable from CM Ethernet connector is plugged in to PC.
If connected, check for dirty or broken contacts at each end.
Current < Min OFF
Press switch once to reset the CM, check cable connections, and
press switch again to retry turning on power.
Current < Normal OFF
Cycle power to ARIS.
Voltage < Min OFF
Check that the external AC/DC 48 V power supply is plugged in and
has the indicator LED on; cycle power to ARIS.
Temperature > Max OFF
Turn off the CM, and let it cool down; move the unit to a cooler
location (e.g. not in direct sunlight) if possible.
Breaker Fault OFF
Turn off CM. Disconnect sonar cable, and turn on CM.
If no breaker fault then turn off CM, connect cable, and turn on CM.
Power > Max OFF
Cycle power to ARIS.
Voltage > Max OFF
Replace the external AC/DC 48V power supply. If not using the
supplied SMC power supply, ensure that 24 V <= Vin <= 48 V.
Current > Max OFF
Cycle power to ARIS.
NOTE: For persistent error codes, contact Sound Metrics