Chapter 4
Troubleshooting
P/N 79000035_SM, Version 1.5 4 - 29 Focus 30/35 Service Manual
Sensor specific messages Subsystem: FI_Angle_VE
Performance
Value Value description Status Ok Counter Error Counter
A
Meas events Ok 1314
Counter of
succesful angle
readings.
0- Ok status
1- Error message
B Meas event timeout Ok - 1- Error message
C Warning flags Ok - 1- Error message
D Error flags Ok - 0- Ok status
1- Error message
Additional results
perfoming a
Diagnostics:
B1 Could NOT read
1000 V angle
readings.
Item System Monitor messages
and message codes stored in
*.fsmx file
Error Log stored in
*.irep file
Recommendation
01 V-angle sensor measurement
failed
V angle sensor reading error
A A1 C D
Drive system Turn event
error warning 6 B D D1
FI: 10 Code: 0xf002
Message: Angle
measurement failed
FI: 25 Code: 0x9c54
Message: Angle
measurement failed
0x9868 brown out reset
dedected
LED illumination (right)
failed.
Exchange Illumination V-
ACB (right).
02 V-angle sensor measurement
failed
V angle sensor reading error
A A1
Drive system Turn event
error warning 6 B D1
FI: 10 Code: 0xf002
Message: Angle
measurement failed
FI: 25 Code: 0x9c54
Message: Angle
measurement failed
0x9868 brown out reset
dedected
LED illumination (left)
failed.
Exchange Illumination V-
ACB (left).