Chapter 6
Adjustment and Verification
P/N 79000035_SM, Version 1.5 6 - 41 Focus 30/35 Service Manual
4 Adjust reticle in
coincidence as
reference.
A. Adjust the
instruments reticle to
the collimators
reticle in
coincidence, see Fig.
6-29 on page 6-40
step 4 and keep
attention to the
erection of the
Reticle, see “Erecting
the Reticle F30” on
page 6-19.
B. For horizontal
shifting turn the
opposite adjustment
screws in the same
direction.
C. For vertical shifting
turn the opposite
adjustment screws in
the same direction.
D. After adjustment in
Hz and V, clamp the
screws to fix the
position. Remove the
reticle illumination
tool.
Tolerance:
CHz < 10cc
CV < 10cc
1 line
width=10cc
Torque value:
20-25 Ncm
2x 1,5 allan key.
2x 1,5 allan key.