St. Jude Medical 3077 User Manual

St. Jude Medical 3077
32 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
5I-17-024X-B-05 © Osypka Medical 2011
MODEL 3077
Single Chamber (SSI)
External Pulse Generator
User’s Manual
Question and Answer IconNeed help?

Do you have a question about the St. Jude Medical 3077 and is the answer not in the manual?

St. Jude Medical 3077 Specifications

General IconGeneral
Model3077
ManufacturerSt. Jude Medical
CategoryMedical Equipment
TypePacing Lead
PolarityBipolar
Length58 cm
ConnectorIS-1
SterilitySterile
FixationActive
MaterialSilicone/Polyurethane

Summary

Preface

General

Read instructions carefully before use and contact customer service for questions.

Checking the Package

Unpack and check for damage or missing items. Inform manufacturer immediately if issues are found.

Writing Conventions of this Manual

Explains conventions used for notes, cautions, and warnings within the manual.

Product Description

Indication

Contraindication

Possible Complications

Precautionary Measures and Warnings

Patient Safety

Electromagnetic Compatibility

Use and Application of MODEL 3077

Design

Details the external pulse generator MODEL 3077, its controls, and terminals, showing front and rear views.

Connecting Temporary Stimulation Leads

Provides step-by-step instructions on how to connect temporary stimulation leads to the MODEL 3077.

Lead Types

Describes different types of temporary pacing leads, including transvenous, heart wires, and permanent leads.

Powering On MODEL 3077

Instructions on how to turn the MODEL 3077 pacemaker on using the Mode switch.

Powering Off MODEL 3077

Instructions on how to turn the MODEL 3077 pacemaker off using the Mode switch.

Determining the Sensing Threshold

Procedure for determining the minimum signal amplitude required to inhibit or trigger the pacemaker.

Determining the Capture Threshold

Procedure for determining the stimulation amplitude needed to capture the heart.

Modes of Operation

Details the different operating modes of the MODEL 3077 pacemaker.

Demand Stimulation (VVI, AAI)

Describes demand pacing where the pacemaker inhibits stimulation upon sensing intrinsic activity.

Asynchronous Stimulation (V00, A00)

Describes asynchronous pacing where stimuli occur at a programmed rate regardless of intrinsic events.

High-Rate Stimulation (x2, x4)

Explains high-rate stimulation for breaking atrial tachycardia by overdriving the patient's heart rate.

Basic Rate, Stimulation Amplitude and Sensitivity Threshold

Details how to adjust basic rate, stimulation amplitude, and sensitivity using rotary dials.

Battery Surveillance

Describes how the pacemaker monitors battery voltage and indicates when replacement is needed.

Device Surveillance

Explains the pacemaker's surveillance circuitry and malfunction indicators.

Environmental and Medical Therapy Hazards

Discusses potential interferences, hazards from medical equipment, and safety precautions.

Storage

Care and Maintenance

Care and Cleaning

Instructions on how to clean and disinfect the MODEL 3077 pacemaker and its accessories.

Changing the Battery

Step-by-step instructions for replacing the 9V battery in the MODEL 3077 pacemaker.

Safety Check-Ups of the Pacemaker

Details regular visual inspections and functional tests required for safe operation of the pacemaker.

Product Return Policy

Information on service, repair, and disposal of the MODEL 3077, including WEEE guidelines.

Customer Service

Technical Data

Delivery Unit

Conformity According to IEC 60601-1-2

Declaration of Conformity

Related product manuals