What to do if the room alarm is triggered but there is no call at the nursing station with Stanley Healthcare Medical Equipment?
- Jjessica95Aug 6, 2025
Several factors can trigger a 'Room alarm triggered - No call at nursing station' or prevent the nurse call from turning off after an alarm with Stanley Healthcare Medical Equipment. First, ensure the nurse call plug, call cord, and adapter are firmly connected to the patient station. Verify that the Control Unit was reset after an alarm before the nurse call system alarm was canceled. The nurse call adapter should be tested for proper functioning and it should be verified that the mat is a Bed-Check[®] Sensormat[®]. Also, check if the control unit is set to the proper nurse call system configuration (N.O. or N.C.). If it is not, the unit will alarm the nurse call system as soon as power is applied to the unit.