OPERATING THE DEVICE
MAINTENANCE
NOTE: All gaskets must be renewed after each assembly. Make sure the device and connected
accessories are depressurized before carrying out any work on the device.
If over time, the pump starts up after a slight removal of water, approximately 0.5L, the preliminary lling
pressure in the boiler must be re-established.
1. Disconnect the mains plug
2. Open the pressure line (turn on the water spigot or nozzle), allow water to completely run off
3. Unscrew the plastic buttery valve on the front side of the boiler; the air-control valve is located
behind it.
4. Attach the air pump or compressor hose to the air-control valve with a “tire valve” connection and
pressure gauge.
4. Pump up, or inate to the designated preliminary lling pressure. Preliminary lling pressure: 1.8 ~
2.0 bar
Unscrew lter housing (A) and take out the lter cartidge
If necessary, soak the cartridge in warm water
Rinse the lter cartidge with clean water, brush out stubborn dirt from inside
the cartridge using a soft brush.
DANGER: If the conveyed water is highly contaminated, clean the water after
each use. A clogged lter may damage the device.
6
STORAGE
If there is danger of frost, dismantle the device and accessories, clean and store them in a place
protected from frost.
1. Turn off the device and unplug the mains plug
2. Open pressure line, allow water to completely run off
3. Completely empty the pump and boiler
4. Dismantle the suction and pressure lines from the device
5. Store the device in a frost free room, at least 5 °C
Dismantling