3. Remove a small specimen collection tray from its packaging.
4. Open the collection tray port door of the specimen collection manifold.
5. Insert the collection tray into the specimen collection manifold.
6. Suction fluid waste and specimen (polyp) material from the surgical site as required.
7. Examine the contents of the illuminated specimen collection manifold tray.
NNoottee -- Collected specimens can be observed by looking through the top viewing window on the specimen collection
manifold.
8. After a specimen is collected and transfer of the specimen is appropriate, remove the collection tray from the manifold
by pinching the tabs of the tray together and pulling the collection tray away from the manifold (Figure 29).
FFiigguurree 2299 –– RReemmoovvee SSppeecciimmeenn CCoolllleeccttiioonn TTrraayy
NNoottee -- To maintain suction after removing the collection tray, insert another collection tray into the specimen collection
manifold immediately or close the collection tray port door of the specimen collection manifold. Suction does not need to be
turned off when inserting or removing a specimen collection tray.
9. Transfer the specimen into the appropriate container using the illuminated work surface on top of the rover.
NNoottee -- The specimen collection tray may again be installed in the specimen collection manifold during the same
procedure. Close the collection tray port using the cap if the specimen collection tray is not used.
10.After the procedure, perform
To Shut Down the Rover
(page 47).
EN 44 700000969828 Rev-AA