Error codes 䫭䇃ҷⷕ
The following error codes may appear on the
display during the course of using the Duo
Instrument. Before returning an instrument,
ensure that the batteries are not low.
E=1. Infra Red communication error - Repeat
operation
.㑶䖲䫭䇃˗䞡ᮄ䖲
E=2. Measurement attempted when not
calibrated. Calibrate the TU
ޚҾ఼
E=3. Traverse Unit - Oscillator fault - If fault is
persistent, return to a Taylor Hobson Service
Centre.
ᤃ㤵఼䫭䇃˗ᓎ䆂䖨ׂ
E=4. Traverse Unit - Motor Positioning fault - If
fault is persistent, return to a Taylor Hobson
Service Centre.
偀䖒ԡ㕂䫭䇃˗ᓎ䆂䖨ׂ
E=5. Traverse Unit - NVRAM Error - If fault is
persistent, return to a Taylor Hobson Service
Centre. NVRAM
䫭䇃˗ᓎ䆂䖨ׂ
E=6. Traverse Unit - Profile Over-range.
Re-attempt measurement.
䕂ᒧ䍙䞣˗䞡䆩
E=7. Traverse Unit - Parameter Over-range.
Re-attempt measurement.
খ᭄䍙䰤˗䞡䆩
E=8. Traverse Unit - Motor Speed Setup error.
Re-attempt Setup procedur
偀䖒䗳ᑺ䆒㕂䫭䇃
E=9. Traverse Unit - Gain Setup error.
Re-attempt Setup procedure.
Ⲟ䫭䇃
E=10. Control Unit - Oscillator fault - If fault is
persistent, return to a Taylor Hobson Service
Centre.
ࠊܗ˖ᤃ㤵఼䫭䇃ˈᓎ䆂䖨ׂ
E=11. Traverse Unit - Calibration information
download error. Try again.
ޚ䫭䇃ˈ䞡䆩
E=12. Traverse Unit - Not set up. Run the
Setup procedure.
偅ࡼܗ˖䆒㕂
E=13. <Reserved>
E=14. : Traverse Unit - Electronic Measure-
ment error - If fault is persistent, return to a
Taylor Hobson Service Centre.
⬉䏃䫭䇃ˈ䖨ׂ
Specification
Traverse length. 5mm (0.2in)偅ࡼ䭓ᑺ
Measurement length. 4mm (0.16in)⌟䞣䭓ᑺ
Traverse speed. 2mm/s (0.08in/s)⌟䞣䗳ᑺ
Gauge range. 200μm (8000μin)䞣
Gauge resolution. 0.05μm (2μin)
ߚ䕼⥛
Display resolution:Ra. 0.01μm (0.4μin)
Display resolution: Rz, Rt*, Rp*, Rv*.
0.1μm (1μin)
ᰒ⼎ߚ䕼⥛
Cut off. 0.8mm (0.03in) +/- 15%
পḋ䭓ᑺ
Uncertainty. 0.1μm (4μin) or 5% of
result, to 95% confidence level
ϡ⹂ᅮᑺ
Results range:Ra. 40μm (1600μin),
see ISO 4288-1996
᳔5Dؐ
Results range: Rz, Rt*, Rp*, Rv*.
199.9μm (8000μin), see ISO 4288-1996
Stylus. Diamond, radius 5μm nominal , 90
o
tip
angle, skid radius: 10mm nominal.
⌟䩜
Filter type: 2CR, (200:1)Ⓒ⊶఼
Sample spacing: 1um䞛ḋ䯈䱨
Stylus Force. 10mN (1gf) max at mid range
⌟
Batteries. Control Unit: One x 3V Lithium
⬉∴
2450. Traverse Unit: Three x 3V Lithium 2450
Units. μm or μinԡࠊ
Operating conditions. 10 to 35
o
C,
80%RH non-condensing
⦃๗㽕∖
Size. 125mm x 80mm x 38mm
(4.92in x 3.15in x 1.5in)
ሎᇌ
Weight. 200g (7oz)䞡䞣
*5 parameter model only.
Surtronic Duo
112-2916 (2 parameter model) and 112-3115 (5 parameter model)
Operator’s Leaflet K505-54
Issue 1.1 January 2002
Leicester, England
www.taylor-hobson.com
Declaration of Conformity
Manufacturer’s Name: Taylor Hobson Limited
Manufacturer’s Address: 2 New Star Road, Leicester, England, LE4 9JQ
It is declared that the product: Surtronic Duo
Conforms to all applicable requirements of BS EN 50081-1:1992, BS EN 50082-1:1998 and
BS EN 61010:2001.
The above product complies with the requirements of the EMC Directive 89/336/EEC as amended.
The above product complies with the requirements of the EMC Directive 73/23/EEC as amended.
Traverse Unit偅ࡼܗ
Control Unitࠊܗ
Parameter Selector
খ᭄䗝ᢽ
Mode selectorᓣ䗝ᢽ
Start buttonsᓔྟᣝ䪂
Stylus
⌟䩜
Mode statusᓣ
Parameter statusখ᭄
Displayᰒ⼎ሣ
On buttonᓔᴎᣝ䬂
IrDA windows㑶にষ
Marketing CHINA:TEL:86 (010) 82951585 82946733 FAX:86 010 58859230