Table 1: Meter-Mode Zero & Gain Adjustments
*exclusive of noise
Table 2: Calibrator-Mode Zero & Gain Adjustments
*exclusive of noise
26
Step
Sensor
Select
U.U.T.
Input
Adjust
U.U.T.
Reading*
1
2
3
4
5
6
7
8
9
10
11
12
13
K
K
0.000mV
54.856mV
P1
P11
32.0±0.1°F
2500.0±0.1°F
Repeat Steps 1 & 2 As Required
J
T
E
N
B
R
S
G
C
D
0.000 mV
0.000 mV
0.000 mV
0.000 mV
1.439 mV
0.000 mV
0.000 mV
2.238 mV
0.000 mV
0.000 mv
P1
P1
P1
P1
P1
P1
P1
P1
P1
P1
32.0±0.1°F
.
32.0±0.1°F
32.0±0.1°F
32.0±0.1°F
1000.0±0.5°F
32.0±0.5°F
32.0±0.5°F
600.0±0.2°F
32.0±0.2°F
32.0±0.2°F
Step
Sensor
Select
U.U.T.
Input
Adjust
DMM
Reading*
1
2
3
4
5
6
7
8
9
10
11
12
13
K
K
32.0°F
2500.0°F
P1
P41
0.000 mV±1µV
54.856 mV±1µV
Repeat Steps 1 & 2 As Required
J
T
E
N
B
R
S
G
C
D
P1
P1
P1
P1
P1
P1
P1
P1
P1
P1
0.000 mV±1µV
0.000 mV±1µV
0.000 mV±1µV
0.000 mV±1µV
1.439 mV±1µV
0.000 mV±1µV
0.000 mV±1µV
2.238 mV±1µV
0.000 mV±1µV
0.000 mV±1µV
32.0°F
32.0°F
32.0°F
32.0°F
1000.0°F
32.0°F
32.0°F
600.0°F
32.0°F
32.0°F