Calibration Adjustments:
NOTE: Do not deviate from the calibration adjustment sequence
that follows. This will ensure that adjustments to be stored in
EEPROM go to the correct locations.
1. Perform meter-mode adjustments and verifications per Table 3 (use Table
5 for RTD-Thermistor models).
2. Hit OPR key (Flashing 'RTD' annunciator turns off).
3. Install J1 jumper.
4. Set unit to CALIB mode, RTD-100, °F.
5. Switch current source to OPERATE mode.
6. Perform calibrator-mode adjustments and verifications per Table 4 (use
Table 6 for RTD-Thermistor models).
7. Turn off unit. Remove tape holding battery.
8. Re-install back cover. Resistance-function calibration is complete.
Table 3: Calibration of Meter-Mode Resistance Functions
(RTD-100/1000)
*exclusive of noise
** adjust as necessary to obtain steps 2,5.
30
Step
Sensor
Select
Unit
Output
Adjust
Verify
DMM
Reading*
1
2
3
4
5
6
7
8
9
RTD-100
RTD-100
RTD-100
RTD-1000
RTD-1000
1,000
1,000
10,000
10,000
390.48
100.00
18.52
3,904.8
1000.0
900.00
1.00
9,000.0
10.0
P1**
P1**
P1
P1
1562.0° ±0.5°F
32.0±0.1°F
-328.0±0.2°F
1562.0°F±0.5°F
32.0+0.1°F
900.00
1.00±0.10
9000.0
10.0±1.0