GENERAL INFORMATION
This manual provides operating instructions and maintenance information for
five temperature instruments. These instruments are high performance cali-
brator-thermometers capable of simulating and measuring a wide-variety of
sensors. In addition, features such as data storage and ramping further
enhance their versatility.
It is recommended that you read this manual thoroughly, especially the sec-
tions on safety, prior to operating these instruments.
COMMON SPECIFICATIONS
THERMOMETER INPUTS: THERMOCOUPLE; Miniature TC jack. RTD, OHMS,
THERMISTOR; T-series instrumentation connector.
CALIBRATOR OUTPUTS: THERMOCOUPLE; Miniature TC jack. RTD, OHMS,
THERMISTOR; T-series instrumentation connector.
THERMOCOUPLE TYPES: K, J, T, E, N, B, R, S, G, C, D.
RTD TYPES: 100Ω platinum (α = 0.00385)
1000Ω platinum ( α= 0.00385)
THERMISTOR TYPE: 2252Ω (YSI Series 400).
RANGES: K; -200°C to 1372°C, -328°F to 2502°F
J; -210°C to 760°C, -346°F to 1400°F
T; -200°C to 400°C, -328°F to 752°F
E; -230°C to 1000°C, -382°F to 1832°F
N; -200°C to 1300°C, -328°F to 2372°F
B; 500°C to 1820°C, 932°F to 3308°F
R; 0°C to 1768°C, 32°F to 3214°F
S; 0°C to 1768°C, 32°F to 3214°F
G; 300°C to 2316°C, 572°F to 4201 °F
C; 0°C to 2316°C, 32°F to 4201 °F
D; 0°C to 2316°C, 32°F to 4201 °F
RTD; -200°C to 850°C, -328°F to 1562°F
THERMISTOR; -40°C to 150°C, -40°F to 302°F
OHMS; 0-999.99Ω
0-9,999.9Ω
0-99,999Ω
ACCURACY (18°C to 28°C ambient, 2 years)*:
K, J, T, E, N; ± 0.5°F (rdg ≥ -50°F)
±1.0°F (rdg < -50°F)
B, R, S, G, C, D; ±1.7°F