EasyManua.ls Logo

Tektronix 544 - HORIZONTAL AMPLIFIER

Tektronix 544
126 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
HORIZONTAL AMPLIFIER
WAVEFORM AND VOLTAGE READINGS were obtained under the fol
lowing conditions:
TIME BASE
TRIGGERING LEVEL
............................
+ 4 5 °; pushed in
TRIGGERING:
M O D E .................................................. TRIG
SLOPE.................................................. +
COUPLING
........................................
AC
SOURCE ............................................. EXT
TIME/CM
..................................................
1 mSEC/CM
VARIABLE (TIME/CM)
..........................
CALIBRATED
SWEEP MAGNIFIER............................... XI
HORIZONTAL POSITION (VERNIER) . . ccw
INTENSITY
Normal
WAVEFORM CONDITIONS
TRIGGER INPUT Signal . . .
AMPLITUDE CALIBRATOR . .
HORIZONTAL DISPLAY
___
SWEEP MAGNIFIER
..............
HORIZONTAL POSITION . .
10 Volts P-P Calibrator
10 Volts
NORMAL
XI
Centered
QUIESCENT VOLTAGE CONDITIONS
HORIZ INPUT S ignal
............................
None
HORIZONTAL DISPLAY........................ Ext
HORIZONTAL POSITION:
Upper Voltage Readings................. ccw
Lower Voltage Readings................. cw
Also see IMPORTANT note on Block Diagram.