Application Exa
mples
3. Select Display > Display Persistence >
Innite Persistence.Inthisexample,
you are lookin
g at a clock signal. After
observing the signal for 1–2 minutes,
but before you look for the problem
elsewhere, g
otostep4.
4. Push FastAcq.
MSO70000/C, DSA70000B/C , DPO70000B/C, and DPO7000 Series
MSO5000 and DPO5000 Series
5. Find glitches, transients, or other random
anomalies that are present in the signal.
In thi
s example, FastAcq revealed an
300 ns positive glitch after only a few
seconds.
160 MSO70000/C, DPO/DSA70000B/C, DPO7000, and MSO/DP O5000 Series U ser M anual