Loading...
Step response settling time
≥4 GHz, ≤20 GHz models
The time by which the step response enters and stays below the indicated % error. Step transition
occurs at the 50% amplitude point of the step leading edge.
DSP off
Instrument Gain setting (FS) ± Step
amplitude
Settling
Error
Time
MSO/
DPO72004C
MSO/
DPO71604C
MSO/
DPO71254C
100 mV, 200 mV, 500 mV, 1 V,
2 V, 5 V
≤ 100% FS,
≤ 50% FS
overdrive
< 9%
< 5.5%
< 2.5%
< 0.15%
150 ps
400 ps
3 ns
1 ms
MSO/
DPO70804C
MSO/
DPO70604C
MSO/
DPO70404C
100 mV, 200 mV, 500 mV, 1 V,
2 V, 5 V
≤ 100% FS,
≤ 50% FS
overdrive
< 6%
< 3%
<2.5%
<0.15%
150 ps
600 ps
3 ns
1 ms
DSP on
Instrument Gain setting (FS) ± Step
amplitude
Settling Error
Amount Time
MSO/
DPO72004C
MSO/
DPO71604C
MSO/
DPO71254C
100 mV
200 mV
500 mV
1 V
2 V
5 V
≤ 100% FS,
≤ 50% FS
overdrive
<6%
<3%
<2.5%
<0.15%
100 ps
400 ps
3 ns
1 ms
MSO/
DPO70804C
100 mV
200 mV
500 mV
1 V
2 V
5 V
≤ 100% FS,
≤ 50% FS
overdrive
< 6%
< 3%
< 2.5%
< 0.15%
150 ps
600 ps
3 ns
1 ms
MSO/
DPO70604C
100 mV
200 mV
500 mV
1 V
2 V
5 V
≤ 100% FS,
≤ 50% FS
overdrive
< 6%
< 3%
< 2.5%
< 0.15%
200 ps
800 ps
3 ns
1 ms
MSO/
DPO70404C
100 mV
200 mV
500 mV
1 V
2 V
5 V
≤ 100% FS,
≤ 50% FS
overdrive
< 6%
< 3%
< 2.5%
< 0.15%
300 ps
1.2 ns
3 ns
1 ms
Vertical system analog channels (cont.) Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C
series)
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series 15