Command Descriptions
2–270
TDS Family Oscilloscope Programmer Manual
 specifies one of the input channels.
 
selects channel 1 as the input source for the delayed trigger.
TRIGger:DELay:EVENTS? (Query Only)
Returns the current delayed trigger event parameter.
Trigger
 




might return 
TRIGger:DELay:EVENTS:COUNt
Sets or queries the number of events that must occur before the delayed trigger
occurs when TRIGger:DELay:BY is set to EVENTS. This is equivalent to
setting the Delay by Events count in the Delayed Edge Delay side menu.
Trigger
  
 






 is the number of delayed edge trigger events. The TDS 400A range is 1 to
(10E7 –1). The TDS 510A, 500C, 600B, and 700C range is 2 to 10E7.