Command Descriptions
TDS200, TDS1000/2000, TDS1000B/2000B, TPS2000 Programmer
2- 107
ALLEv?, EVENT?, EVMsg?
FACtory (Set Only)
Resets the oscilloscope to its factory default settings. Refer to
Appendix B: Factory Setup for a list of the factory default settings.
Miscellaneous
FACtory
Setting the oscilloscope to factory default has the following impact
on the programming interface:
H Clears the Event Status Enable Register
H Clears the Service Request Enable Register
H Sets the Device Event Status Enable Register to 255
H Sets the Power On Status Clear Flag to TRUE
H Enables all Command Headers (HEADer ON)
H Sets the macro defined by *DDT to a “zero-length field”
H Clears the pending operation flag and associated operations
H Performs the equivalent of DATA INIT
The FACtory command does not alter the following items:
H State of the RS-232, GPIB, or USB int erface
H Selected GPIB address
H Front panel LOCK state
H Verbose state
H Display Brightne ss, TPS2000 serie s only
H Display Contrast, TDS1000, TDS2000, TDS1000B, TDS2000B,
and TPS2000 serie s