CENtronics specifies that the hard copy data is sent out the
Centronics port.
RS232 specifies that the hard copy data is sent out the RS232 port. If
you set the port to RS232, a nd use it to t ransfer a BMP screen image
format file to a PC or other computer, observe the following
precaution that the BMP file is a binary file, and therefore does not
use linefeeds (hexadecimal 0a) as a terminator.
There a re two ways to work around this limitation:
H Write your controller program to use the byte count inform ation
in the BMP file header.
H Set your RS232 program’s time out value to a large enough value
to guarantee that the program does not time out before complet-
ing the data transfer. For example, to transfer an 80 kB file at
9600 baud, set the transfer program time out to 300 seconds.
GPIb specifies that the hard copy data is sent to the GPIB port. This
argument is available on instruments with a TDS2CM, TDSCMA,
TDSCMAX, or TDSMM module insta lled.
USB specifies that the hard copy data is sent to the USB Device port
on the back of a TDS1000B or TDS2000B series oscilloscope only.
HARDCopy:PORT?
Might return RS232 as the sele cted hard copy output port.