Measuring Waveforms
CSA7000 Series, TDS7000 Series, & TDS6000 S eries Instruments User Manual
3- 183
Overview Related control elements and resourcesTo deskew channels (Cont.)
Compensate
probe timing
(deskew)
(Cont.)
12. Touch Deskew Time and use the mul tipurpose knobs or
keypad to adjust the deskew time for that channel so
that its signal aligns with the trigger position.
13. Repeat steps 11 and 12 for each additional channel that
you want to deskew.
14. Remove the connections.
For further
assistance
15. Touch the Help button to access the online assistance.
See page 3--283 to learn about using online help.