EasyManua.ls Logo

TEL AFA1000 / 2 – Mk2 - Page 9

TEL AFA1000 / 2 – Mk2
17 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9
High / Low
When input configured as High/Low is activated
Display Icon shows High or Low
High / Low relay operates ( if configured )
This function is designed for two speed fan operation or two position damper operation
switched via a micro switch or proximity switch activated at a given position on the sash.
Night set-back
When input configured as Night set-back is activated
Night Set-Back Icon is displayed
Red LED on ( Flashing)
Reduced Low air alarm ( if configured )
Audible alarm muted
Mute Icon shown on display
External alarm
When input configured as External alarm is activated
Red LED on ( Flashing) – ( if configured )
External Alarm toggles on /off with display -- ( if configured )
Audible alarm sounds – can be muted via Enter pushbutton
External alarm relay operates ( if configured)
Alarm disable
When input configured as Alarm disable is activated
Alarm disabled is displayed
Red LED on ( Flashing)
Audible alarm muted
Mute Icon shown on display