Physical Characteristics:
Size: 17.3 cm (H*) x 14.0 cm (W) x 20.6 cm (D)
6.8” (H*) x 5.5” (W) x 8.1” (D)
Weight: 2.8 pounds/1.3 Kg (approximate)
Note: Allow 33 cm (13”) total height allowance for column installation
T
ransportation and Storage:
Temperature: -20 to +50 °C (-4 to +122 °F)
Relative Humidity: 95% max (non-condensing, Transportation)
85% max (non-condensing, Storage)
Electromagnetic Comp
atibility (EMC):
This device meets the requirements of IEC and UL 60601-1 Standards for Medical Electrical
Devices with respect to Electromagnetic Emissions and Compatibility. This device can be placed
in close proximity to other Medical Devices approved to the same standards. Communication
devices or other equipment emitting RF signals may cause erratic operation of this device.
(See Tables, Appendix I)
S
tandards:
The ConchaTherm Neptune Heated Humidifier will be tested to and will comply with the following
standards:
UL 60601-1, Medical Electrical Equipment, Part 1: General Requirements for Safety
IEC 60601-1 Medical Electrical Equipment
ISO 8185 2nd Edition; Humidifiers for Medical Use – General Requirements
or Humidification Systems
IEC 60601-1 Clause 5 Classifications:
• Class I
• Type BF
• IPX1
• Rated of Continuous Operations