LED flashes red.
Sealing unit is too hot.
Turn the power switch off and wait about
10 minutes, and then try again.
If the LED is still flashing, contact your
local Terumo BCT representative.
LED doesn’t light.
Sealing unit defective.
Contact your local Terumo BCT
representative.
Sealing does not start.
Wet electrodes.
Leaky or wet tube.
Blocked optic sensor.
Other cause.
Clean the electrodes as described in
Section 6.1.
Dry the tube and try again.
Clean the optic sensor.
Contact your local Terumo BCT
representative.
Tube splits or breaks during
sealing.
Tube stretched during sealing.
Seals are too close together.
Incorrect adjustment of
electrodes.
Do not stretch the tube. See Section 5.2.
Leave a minimum of 1 cm (0.39 in)
between seals.
Contact your local Terumo BCT
representative.
8. Warranty and Service
If any serious incidents occur from the use of this equipment, please report the event, as soon as possible, to the closest
Terumo office.
Information on the warranty and the service provided by Terumo BCT is listed below.
8.1 Warranty
Terumo BCT guarantees that the equipment shall be free from defects in material and workmanship when delivered to
the original purchaser. Terumo BCT´s sole obligation shall be limited to repair or replacement, at Terumo BCT´s option
and expense, of the defective part or unit for the duration of the warranty.
The warranty extends only to the original purchaser and is not assignable or transferable and shall not apply to auxiliary
equipment or disposable accessories.
Terumo BCT guarantees that the equipment is fit for the purposes and indications described in the labelling when used
in accordance with the instruction for use. Unless the equipment is used in accordance with such instructions, this
warranty is void and of no effect. No other expressed or implied warranty exists, including any warranty of
merchantability or appropriateness for a particular purpose. Terumo BCT’s sole obligation and original purchaser’s
exclusive remedy for breach of warranty shall be limited to repair or replacement at Terumo BCT´s option.
Terumo BCT shall not be liable for proximate, incidental or consequential damages. Modifications, alterations,
recalibrations or abuse, and service by other than a Terumo BCT authorized representative will void the warranty.
8.2 Service
Service Under the Warranty Period
While under Terumo BCT warranty, the instrument must not be opened by unauthorized personnel.
Contact your Terumo BCT representative for service and repair information for all T-SEAL III instruments.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Terumo T-SEAL III and is the answer not in the manual?

Terumo T-SEAL III Specifications

General IconGeneral
BrandTerumo
ModelT-SEAL III
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals